The brand new Data Output Module is the newest addition to the flexible multi-gas system. The Data Output Module is an accessory for the AX60+ fixed gas monitor. This module has two features. The first, a 4-20mA signal over 4 channels with one per sensor. The second, a MODBUS RTU Communications.

Do you use an AX60 in your facility? You need to be aware that it is sensitive electronics. Warning notifications need to be issued to all staff members. They must also be placed around your building. Please contact one of our team for any questions: 01924 444577 or hello@gemscientific.co.uk

Description

Analox Data Output Module For AX60+

The brand new Data Output Module is the newest addition to the flexible multi-gas system. The Data Output Module is an accessory for the AX60+ fixed gas monitor. This module provides the user with real-time monitoring of the AX60+. Whilst also communicating with customers remote systems such as a Building Management System (BMS). This module has two features. The first, a 4-20mA signal over 4 channels with one per sensor. The second, a MODBUS RTU Communications. Both features are able to communicate a sensor measured value and status to a master control panel or screen in a different location. This makes it easier for the user to monitor all reading from devices in their facility in one location. The Data Output Module is compatible with any Analox AX60+ including previously installed AX60+ systems (except the AX60 and AX60+ kiosk systems). This module can also be retrofitted to any existing AX60+ installations.

4-20mA

The 4-20mA is an industry standard communication method for communicating sensor status to remote locations. This unit is easy and simple to install.

Features

  • Fault diagnostics (e.g. sensor is in fault)
  • 4 channels per 1 sensor
  • Transmits gas levels (e.g. 20.9% O2)

MODBUS

The MODBUS enables more detailed communications regarding sensor status and it allows monitoring of the AX60+ system for more in-depth analysis. The Data Output Module is simple and easy to mount and install. The Data Output Module can be attached to a wall or suitable flat surface. The user can also connect into the AX60+ using the supplied cable. This unit is available as a hard-wired unit or as a quick connect

 Features

  • More detailed communications
  • Depth analysis
  • Transmits gas levels (e.g. 5000ppm CO2)
  • Transmits alarm status (e.g. AL2 triggered)
  • Ability to transmit fault status (e.g. sensor 2 is in fault)
  • Enables the transmission of multiple data messages from each sensor

Guide to Gas Monitoring in the Hospitality Industry

Specifications  D.O.M AX60+
Number of analogue outputs 4
Output load capacity 500 ohms max loop resistance per output
Analogue output type Active 4-20mA per output
Digital output RS-485
Comms protocol Modbus RTU
Mounting Wall mounted as per sensor
Certification CE & IEC 61010-1 (UL) plus Australian AS5034-2005
Warranty 2 years
Dimensions (WxHxD) 72 X 132 X 52